|
In
order to estimate organ dose, we need first be able to accurately quantify
organ activity at one time point. In
vivo organ activity estimations are usually done by conventional planar
(C-Planar) processing methods which are based on conjugate-view whole body
scans and geometric mean (GM) method. Figure 1 below shows typical steps in a
C-Planar method. This method is based on acquisition of anterior and
posterior images that are scatter corrected using the triple energy window
method. Attenuation compensation is the performed using the geometric mean
method using thickness correction with a measured transmission image. ROIs are then drawn on the planar image. Since organs
overlap in the 2D projections, and since it is hard to see the true organ
outlines in the 2D projections, ROIs are often
drawn smaller than the true projected organ size and, in some cases,
additional corrections for background and organ overlap is performed. The
last step is the use of a calibration factor to convert counts in the ROI to
total organ activity.

Figure
1. Typical steps of C-Planar method
There
are some obvious shortcomings of the C-Planar method. First, the GM method
only approximately compensates for attenuation. Also, the triple energy window
method is only approximate, results in an increase in image noise, and may
require more energy windows than are available on commercial gamma cameras.
Also, there are no compensations for collimator-detector-response or partial
volume effects, which are very important for tumor dosimetry.
Perhaps the major problem is organ overlap and background activity, as
illustrated in Figure 2.

Figure
2. Overlapping problems in the C-Planar method
The
ROIs on the leftmost image are the 2D projection of
the 3D organ VOIs defined in co-registered SPECT
and CT images. We can see that the organs overlap in the 2D projection view.
In C-Planar methods, smaller ROIs are often used to
reduce the overlap and the impact of background activity. However, this will
result in leaving out activity in some portions of the organ. In some
methods, the missing activity is extrapolated based on the activity in the
non-overlapping portion and background activity is estimated using background
ROIs. However, the entire process is very subjective
and it is difficult to know a priori, for example, which set of ROIs will provide more accurate activity estimates. For
example, the ROIs in the rightmost image will
drastically reduce the effects of background and overlap, but result in
leaving out activity larger portions of each organ.
One
potential solution to the overlap and background problem is the use of SPECT
images, as demonstrated in Figure 3. Since we can define organ volumes of
interest in 3D, the overlap problem is largely eliminated. However, SPECT
images without compensation for physical effects are not quantitative. Even
with compensation, the limited resolution will result in partial volume
effects that will degrade quantitative accuracy. We have developed and are
continuing to develop quantitative SPECT (Q-SPECT) methods in our lab that
provide comprehensive compensation for these effects. We have applied,
refined, and evaluated these methods for organ activity estimation.

Figure
3. The three leftmost columns show transaxial views
of SPECT,
CT and fused SPECT/CT images,
respectively, overlaid with the manually-defined organ ROIs
for the corresponding slice. The rightmost 3 columns show coronal (top row)
and sagittal (bottom row) slices from the (left to
right) SPECT,
CT, and fused SPECT/CT images.
In
the Q-SPECT methods, we used the OS-EM algorithm with the compensation for
attenuation, scatter, and the full collimator-detector-response, which
includes modeling penetration and scatter in the collimator. A
perturbation-based geometric transfer matrix (pGTM)
method developed in our lab was used for partial volume compensation. Organ
activities were then calculated from the total reconstructed image intensity
divided by the system sensitivity.
Although,
Q-SPECT can provide very accurate activity estimates, there are also some
limitations on using it clinically. First, it requires more complex imaging
protocols. For example, it requires 2 scans to cover all the important organs
without the whole-body SPECT capacity. However, this may become a less
important factor in the future if spiral scan whole-body SPECT protocols are
developed. The acquisition time is normally longer than WB scans, though in
other imaging applications there is some data to suggest that good image
quality can be obtained with the same acquisition time as for the planar
scans. Also, the computation time is much longer than for QSPECT methods and
the definition of the 3D VOIs is tedious since
robust and general automatic segmentation tools are not available. However,
the availability of co-registered SPECT/CT does ease and improve the accuracy
of image segmentation.
To
address some of the limitations of Q-SPECT, we have developed a quantitative
planar method we will refer to as Q-Planar. As demonstrated in Figure 4, we
still use the conjugate whole body scans, but using an iterative algorithm to
estimate the organ activities from these two projections in Q-Planar method.
As shown here, we still make use of 3D information from CT or SPECT. First,
we segment out each organ in 3D, and uniformly fill each VOI so sum of all
the voxels inside the VOI is 1. Then we use our
model-based projector to project the VOI image. The true projection will then
be a linear combination of the separate organ VOI projections. Since the
projections are Poisson distributed, we can use the ML-EM algorithm to
estimate the scale factors, Ar, for the organ VOI projections and these scale factors
will simply be the total activity in the organ. The major difference between
this and SPECT reconstruction is that we need to estimate only a few
parameters and thus can estimate them quite effectively, as we will show,
from only 2 projection views.

Figure
4. Flow chart of activity quantification steps in Q-Planar method. First,
each organ VOI is segmented from the registered 3D SPECT/CT data, and
uniformly filled so that the sum of all the voxels
inside the VOI is 1. Then the projection of each organ VOI is calculated
using the model-based projector. The true projection will then be a linear
combination of the separate organ VOI projection. Since the projections are
Poisson distributed, we can use the ML-EM algorithm to estimate the scale
factors, Ar,
for the organ VOI projections and these scale factors will simply be the
total activity in each organ.
Since
we can estimate the organ activity from both planar and SPECT scans, there
are at least three different acquisition strategies for estimating cumulated
activity or residence time. As listed here:
1.
A time series of conjugate view whole body
scans;
2.
A time series of SPECT scans;
3.
A time series of conjugate view whole body
scans plus a SPECT scan at one time point, which referred as the hybrid
method.
The
residence times can be estimated using a class of methods for each protocol.
Note that the planar acquisition protocol will be,
in general, easier to implement than the SPECT acquisition protocol, and the
hybrid protocol is a practical compromise between the two.


Figure
5. Flow chart of residence time quantification steps in planar, SPECT, and
hybrid planar/SPECT methods.
|
|
We
performed a Monte Carlo (MC) simulation (MCS) study using the 3D NCAT
phantom. The organ activity concentrations were based on the averages from 8
clinical studies using Indium-111 Zevalin. We used
a non-uniform activity distribution in the heart and lungs.
The
simulation parameters were appropriate for a GE VH/Hawkeye camera with a 1
inch crystal and a medium energy general purpose collimator. We used a
modified version of SimSET and PHG code that
includes modeling of collimator interactions to simulate SPECT projections
and planar images. The low-noise projections were generated using this code.
The resulting organ and background projection images were scaled to represent
organ activities at different time points (1, 5, 24, 72 and 144 hours) based
on the average organ time activity curves obtained from 6 patient studies and
summed to form a low-noise set of projections for each time point. Fifty
different Poisson noise realizations were generated to study the precision of
the methods. The SPECT projections were simulated at 120 views over 360
degree, the count level at 24 hour is equivalent to 30 seconds scan per view.
The count level of planar scans at 24 hour is equivalent to 20 minutes whole
body scans.

Figure
6. From left to right the images are: coronal slice through activity
distribution, same coronal slice through attenuation map, low-noise anterior
projection, the noisy anterior SPECT projection, and the noisy anterior
planar projection.
As
mentioned, for the CPlanar method, organ overlap is
an important issue. For the MC study we investigated 3 different cases of
overlap correction, as demonstrated in Figure 7.
1.
For the ideal correction we took advantage of
the fact that all the organs were simulated separately in the MC simulation.
We applied the CPlanar method to the projection of
each organ to estimate the organ activity. Thus there was no overlap with
other organs and no background activity. This is an ideal case and no real
method could do better than this.
2.
At the other extreme is no correction. In this
case we defined the planar ROIs based on
projections of the true 3D organ VOI images, as shown in the topmost image.
This method has the maximum overlap.
3.
In between these two extreme cases, we
investigated what a realistic overlap and background correction based on the
use of manually drawn ROIs. In this case, organs ROIs were intentionally drawn smaller to avoid
overlapping. No additional overlap or background corrections were performed.
As I mentioned earlier, this method is somewhat subjective, and thus we used
the previous two methods to provide brackets for the kind of variability one
might have.

Figure
7. Three different cases of overlap correction in C-Planar Method.
Figure
8 shows the percent errors of activity estimates for C-Planar method with the
3 different overlap corrections, Q-Planar and Q-SPECT. The vertical axis is
the percent error in organ activity estimate, calculated as the estimated
value minus true value divide by true value. A positive error indicates
overestimation. The error bars were calculated from the fifty different noise
realizations. Please note this is precision due to noise, which is measured
using a single phantom over many noise realizations.

Figure
8. Percent errors and standard deviations of errors in organ activity
estimates for C-Planar, Q-Planar, and Q-SPECT methods.
The
results of this study show that C-Planar w/o overlap correction performed worst,
as expected, with errors up to 380%. Even with ideal overlap correction, the
C-Planar method still produced errors in the range of -8% to 12%. Most of
these errors resulted from the approximate scatter and attenuation
compensations. When using manually defined ROIs,
the errors were generally between the other two extreme
cases, and is perhaps indicative of what would be realized clinically.
The precisions for all the methods were similar and much smaller than the
errors, indicating that noise is not as important a factor as bias from
imaging and processing methods. The results also showed that Q-Planar was
significantly better than realistic C-Planar, with the accuracy approaching
that for Q-SPECT and with slightly better precision.
Figure
9 shows the percent errors of residence time estimates for C-Planar,
Q-Planar, Q-SPECT, and hybrid planar/SPECT methods. Similar to the errors in
activity estimates, the results show that C-Planar w/o overlap correction
performed worst, with errors up to 302%. Even with ideal overlap correction,
the C-Planar method still produced errors in the range of -8% to 7%. When
using manually defined ROIs, the results was
somewhat between the other two extreme cases. The hybrid C-Planar (Realistic)
with QSPECT method performed better than C-Planar methods alone. Q-Planar or
hybrid Q-Planar/Q-SPECT was significantly better than C-Planar methods, with
the accuracy approaching that for Q-SPECT.

Figure
9. Percent errors and standard deviations of errors in residence time
estimates for C-Planar, Q-Planar, Q-SPECT, and hybrid planar/SPECT methods.
|